9 Technical Specifications
_____________________________________________________________________________________________________________________________________
9-13
9.5 AC Current
CONVERSION METHOD.... True RMS value with DC Current coupling or
pure AC Current
RANGES ............................... 200µA / 2mA / 20mA / 200mA / 2A
RANGE SELECTION........... manual, automatic
MEASUREMENT TIMES Full Scale max. Resolution
100ms 201 000 1nA
0.2s - 100 s 2 010 000 100pA
Range 2A 2 010 000 1µA
ACCURACY (1 Year, 23°C ± 5°C) 1)
± (% of Reading + % of Full Scale)
Range 20Hz to 40Hz to 1kHz to 5kHz
200µA 0.2 + 0.01 / 0.08 + 0.05 / 0.2 + 0.05
2mA 0.2 + 0.01 / 0.08 + 0.05 / 0.1 + 0.05
20mA 0.2 + 0.01 / 0.08 + 0.05 / 0.1 + 0.05
200mA 0.2 + 0.01 / 0.08 + 0.05 / 0.1 + 0.05
2A 2) 0.2 + 0.01 / 0.08 + 0.05 / 0.4 + 0.05
DC Current ± (0.15% Rdg. + 0.2% F.S.)
TEMPERATURE COEFFICIENT
(10°C-18°C and 28°C - 40°C) ±(0.005%Rdg. + 0.001% F.S.)/°C
(0°C-10°C and 40°C - 50°C) ±(0.010%Rdg. + 0.002% F.S.)/°C
BURDEN VOLTAGE
Range
200µA 250mV
2mA, 20mA 250mV
200mA 250mV
2A 600mV