CHANNELS ............................. 80 Channels 1-pole, 40 Ch 2-pole, 20 Ch 4-pole
SWITCHING METHOD ......... bi-stable mechanical relays
MEAS. FUNCTIONS .............. Voltage, Resistance, Temp., Hz
THERMO VOLTAGE ........... typ. ± 1µV, max. 2µV after 1.5h warm-up time
MAX. VOLTAGE BETWEEN
2 CONTACTS .................... 125 Vpk with the limitation of 100 000 * V * Hz.
MAX. MEAS. VOLTAGE ....... 125Vpk (also over the V/Ω-Connector) with the
limitation of 100 000 * V * Hz.
MAX. CONTINUITY RESISTANCE
(PER LEAD) .................. approx. 1 Ω
LIFETIME ................... 2 * 100 000 000 Alternations (0.1 A, 10 Vdc)
ISOLATION RESISTANCE BETWEEN
2 CONTACTS ......................... 3 GΩ at relative ambient humidity under 60%
ISOLATION RESISTANCE AGAINST
CASE ............................. 3 GΩ at relative ambient humidity under 60%
CAPACITANCE ....................... smaller than 100 pF between the contacts
TIME BETWEEN TWO SWITCHING ACTIONS 20ms
DELAY BEFORE MEASUREMENT START AFTER
CHANNEL SWITCH.... depends on Function and Range
between 100ms and 500ms
ACTIVATE SCANNER with internal socket (see Ch. „Getting Started“)