EasyManua.ls Logo

Prema DMM 5017 - PIN Number and Calibration Switch

Prema DMM 5017
153 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PIN Number and Calibration Switch
_____________________________________________________________________________________________________________________________________
6-4
6.6 PIN Number and Calibration Switch
Calibration can be protected against accidental or incorrect calibrations, through the
use of a rear calibration switch and a special PIN number.
The calibration button is located on the rear panel, and can be pushed with a pointed
object (pen, pencil, paper clip, etc.). This button can be pushed after the calibration
menu has been selected. At the same time, a request to input the PIN number will
appear in the display.
Press Key Display / Action
1. Cal key on the
rear panel
1 PIN:0 000000
Enter a PIN No. (after delivery: 0000000), the display shows alternating
"CAL" and the reading with the correct PIN No..
2. MENU-IN KEY 2 CONFIGURE..
if this display does not appear press the ñ KEY until it appears
3. MENU-IN KEY 4 Calibration..
if this display does not appear press the ñ KEY until it appears
4. MENU-IN KEY 1 Enter Value..
only with the right PIN No..
5. MENU-IN KEY V +100.00000
Enter the calibration value, close with the MENU-IN KEY
6. MENU-IN KEY 1 Enter Value..
press several times to return to the calibration input menu
7. ò KEY 2 Store Cal.
with afterwards pressing of the MENU-IN KEY stores durably the
calibration data
8. Cal key on the
rear panel
to close the calibration procedure
Note: Item 7 is very important. If you forget to store the user calibration it is
completely deleted after you switch off the instrument. Then only the
factory calibration is valid.

Table of Contents