3.8 Continuity Test
The continuity test can be selected with the
CONT KEY
.
Continuity Test Activate loudspeaker at 50
Display: open or close
3.9 Selecting Measurement Ranges
The measurement range can be changed with the
Ø key and the × key. The 2
ND
-
AUTO KEY
key is used to automatically pre-select a measurement range (2nd +
MenuOut key). The “AUTO” indication then appears in the status display.
If the measurement range is set manually, and a measurement signal is too large for
that range, “Overflow” will be indicated on the display.
If automatic ranging is on, a higher measurement range is automatically selected
when a display of about 190 000 *) digits is reached. A lower measurement range is
selected when a reading of lower than about 10 000 *) is reached.
Automatic range switching occurs within <5ms.
Note: The measurement range and the integration time are stored for every
measurement function.
*) for 5-½ digit display