Setting the Integration Time / Resolution
_____________________________________________________________________________________________________________________________________
3-8
3.10 Setting the Integration Time / Resolution
The integration time (and consequently the resolution) can be changed with the
TIME
KEY
and the cursor keys. The ×
KEY
increases the integration time, the Ø
KEY
de-
creases the integration time.
Integration Time Resolution
20ms, 40ms, 100ms 5½ Digits
200ms, 400ms, 1s 6½ Digits
2s to 100s 7½ Digits *)
*) for Vdc and
Ω
Table: Integration Times and Resolutions
Note: Range and Integration Time are stored for the respective measurement
function.