5 Remote Control
_____________________________________________________________________________________________________________________________________
5-11
5.7 Operation as Listener
The instrument must be addressed as LISTENER to prepare it for receiving
commands. The instructions for doing this are contained in the user manual of the
computer manufacturer. The segment ‘ADR’ in the right window of the display is lit
when the instrument has been addressed as LISTENER.
The instrument understands the following commands:
MR Measuring mode. The measurement reading is selected and appears in the
display of the digital multimeter and in the character string of the TALKER
function.
CR Calculating mode. The calculation result of the selected program appears in
the display of the digital multimeter and in the character string of the
TALKER function.
Cx Entry of constants,
Format x: Constant No., Sign, Constant value
x = 0: Constant a, x = 1: Constant b Example: C1+3.45678912E+6
Constants has to be entered in the displayed unit, not in the physical unit.
VD Direct voltage Vdc
VA Alternating voltage Vac
VC Alternating voltage + Direct voltage component Vac + dc
O2 2-wire resistance measurement Ω2W
O4 4-wire resistance measurement Ω4W
ID Direct current Idc
IA Alternating current Iac
IC Alternating current + Direct current component Iac + dc
TC Temperature measurement in °C
TF Temperature measurement in °F