Temperature
_____________________________________________________________________________________________________________________________________
9-14
SETTLING TIME................... 100ms at 0.1%
MEASUREMENT PAUSES... after Range, Function, or Channel switching,
300ms
OVERLOAD LIMITS ............ input protection
(3A fast blow out fuse)
Personal Safety:
A-Hi to Ground max. 60Vpk
A-Lo to Ground max. 60Vpk
1) Sine Signal larger than 5% of the Full Scale; V//LO connected to Ground.
2) the specs are to be multiplied with factor 3 for currents > 1A
9.6 Temperature
MEASUREMENT METHOD........ 4-pole Resistance Measurement with
Linearization as per EN 60751 for
Pt10, Pt25, Pt100, Pt500, Pt1000
TEMPERATURE SENSORS...... Pt10 , Pt25, Pt100, Pt500, Pt1000
Platinum Resistance Sensors
DISPLAY RANGE ...... Full Scale Resolution
from 1s
Celsius - 200°C to +850°C 0.001°C
Fahrenheit - 328°F to +1562°F 0.001°F
Kelvin +73 K to +1123 K 0.001 K
MEASUREMENT CURRENT .... Pt10/Pt25/Pt100/Pt500 1 mA
Pt1000 100 µA