6 Calibration
_____________________________________________________________________________________________________________________________________
6-11
4. TEMP-Function: Extended list of RTDs
Press the following buttons in sequence to select the recently calibrated RTD, e.g.
Pt100 :
Menu/↵ A 3 DEVICE ↵ A 3 Temp. Sensor ↵ A 6 Cal. Pt100 ↵ Menu/OUT.
Cal. Pt100 is new in this list as the USER calibrated version of a Pt100-Sensor. The
DIN-EN 60751 calibrated version is still available in the menu as
A 3 Pt100 but without the addendum Cal.
6.13 Storing Calibration Values
Storage of calibration values is absolutely necessary to store them durably. Every
calibration step is stored only temporarily, so that the files are deleted if you switch of
the instrument.
If you wish to interrupt a calibration procedure and turn off the unit, you can do so, only
if you store the calibration with the menu "Store Cal." before you switch off power.
You could then continue the procedure the next day for example, from the point where
you left off.