5 Remote Control
_____________________________________________________________________________________________________________________________________
5-19
Meaning of the Transmitted Characters
Position (first, last character) of the device message
( 1, 1) "+" positive sign of the mantissa
"-" negative sign of the mantissa
( 2, 10) "x" 8-digit mantissa or text message,
Numerical range ".00000000 - 99999999"
(11, 11) "E" Designates the exponent
(12, 12) "+" positive sign of the exponent
"-" negative sign of the exponent
(13, 13) "x" Magnitude of the exponent
(14, 15) "MR" Measurement reading is output
"CR" Calculation result is output
"Cx" Constant No. x is output
(16, 17) "VD" Direct voltage measurement
"VA" Alternating voltage measurement
"VC" Alternating voltage with direct voltage component
"ID" Direct current measurement
"IA" Alternating current measurement
"IC" Alternating current measurement with direct current component
"O2" Resistance measurement, two-wire configuration
"O4" Resistance measurement, four-wire configuration
"TC" Temperature measurement, display in °C
"TF" Temperature measurement, display in °Fahrenheit
"TK" Temperature measurement, display in Kelvin
"FQ" Frequency measurement
"PD" Period duration measurement
"CO" Continuity check