Error Messages
_____________________________________________________________________________________________________________________________________
5-22
5.12 Error Messages
Error 01 Overflow Measuring range overflow
Sensor? Resistance of the sensor overshoots/undershoots the
defined range
Error 02 Div/0 Division by zero in a mathematical function
Error 03 Br. wires Open source line in 4-wire resistance measurement
Error 04 Offset too high Offset too large after pressing the zero key (302)
Error 05 Cal. Error Calibration error (341-343)
Error 07 Polarity? for RTD or Ω, sense or source wrong polarity
Error 10 not mathematics Not for mathematics (303)
Error 11 not temperature Not for temperature measurement (304)
Error 12 Scanner Mode? Scanner mode unsuitable (305)
Error 13 Front Input Only Only at front sockets (e.g. current measurement)
(306)
Error 14 invalid PIN Invalid PIN number for calibration (307)
Values in brackets (only error message 04 to 14) appear when reading out the error
queue. Error message 01 to 03 are send instead of the measurement value.