6 Calibration
_____________________________________________________________________________________________________________________________________
6-5
By following the steps described above, each measurement function and measurement
range is calibrated.
The following measurement functions and ranges must be calibrated:
Measurement Function Measurement Ranges
Vdc 300mV, 3V, 30V, 300V, 1000V
Vac 200mV, 2V, 20V, 200V, 1000V
Idc 200µA, 2mA, 20mA, 200mA, 2A
Iac 200µA, 2mA, 20mA, 200mA, 2A
Ω4W 300Ω, 3kΩ, 30kΩ, 300kΩ, 3MΩ, 30MΩ
Ω2W 300kΩ, 3MΩ, 30MΩ
Table: Measurement Functions to be calibrated
Vac+dc, Iac+dc, Frequency, Period and Temperature Measurement are calibrated
through the measurement functions named above.
The lower resistance ranges are only calibrated in the 4-wire resistance measurement.
Changing the PIN Number
By default, the 7-digit PIN No. is set to 0000000. If you want to change this number
please proceed as follows.
Press Key Display / Action
1. cal key on the rear
panel
1 PIN:0 000000
Enter a PIN No. (default: 0000000), with correct PIN No. the display
shows aternate "CAL" and the reading.
2. cal key on the rear
panel
to close the calibration
3. cal key on the rear
panel
1 PIN:0 000000
Now you can enter the desired 7 digit PIN No.
Table: Changing the PIN number