Compatibility
_____________________________________________________________________________________________________________________________________
5-26
The Statusbyte
BitNo. 5017 6001 6047 remarks
0 EOC EOC EOC Bit-compatible, EOC = End of Conversion
1 Error - -
2EQ - -
3 - Error Error Bit-compatible
4 MAV - - incompatible
5 ESB - - incompatible
6 RQS RQS RQS Bit-compatible
7 Button Button Button Bit-compatible
Bit compatibility results only on querying solely the respective bit.
The decimal value of the status byte is incompatible.
Selecting function, range and integration time
The range and the integration time are stored separately for every function, function
range and integration time are stored for every channel.
IEEE commands to select the DMM type
‘dmm6001’ to select 6001 / 5001 / 4001 command set
‘dmm6047’ to select 6048 / 6047 command set
‘dmm5017’ to select 5017 command set