Fig. 6: Fitting the threaded connectors
1. Blanking plug
2. Union nut
3. Multiple sealing insert
4. Threaded connector
5. Locking nut
4. Screw the appropriate threaded connectors (4) using the cor‐
responding locking nuts (5) and tighten firmly.
5. Insert multiple sealing inserts (3) into the threaded connec‐
tors depending on the cable diameter used.
6. Guide the cables into the threaded connectors.
7. Connect the cables as shown on the wiring diagram.
8. Tighten the union nuts (2) on the threaded connectors so that
they are properly sealed.
9. Place the front part of the housing onto the rear part.
10. Manually tighten the four housing screws.
11. Once again check that the seal is seated properly. Degree of
protection IP 65 can only be achieved if it is correctly fitted.
IP 65 degree of protection
Assembly and Installation
17