116©2006 Sunrise Medical
SEATING SECTION
In order to eliminate the re-home switch (previously
found on the Smart Seat), a separate Counter Reset
Switch (CRS) is installed that allows the system to “re-
home” during normal operation. This switch is located
on the left side of the wheelchair next to the Recline
pivot point (Fig 3.127).
In order to remove the Recline CRS switch, the shroud
must fi rst be removed. Refer to the Recline Actuator
Removal section in this tech manual.
Complete the following steps to remove and replace the
Recline CRS.
The Recline CRS located near the pivot point is 1.
non-adjustable.
To remove the CRS, use a small Phillips head 2.
screw driver and remove the two screws securing
the switch to the pivot point (Fig 3.128).
Follow the wires up the back until two spade lugs 3.
are located. Disconnect the spade lugs and re-
move the switch (Fig 3.129).
Reverse the process to install a new Recline CRS. 4.
The part number for the CRS is 103953.
Fig 3.127
Fig 3.128
Fig 3.129
G
Recline CRS Replacement