RHYTHM SERVICE MANUAL
PAGE 8.11
2009SUNRISE MEDICAL
Section 8
Adjusting Power ELR / ALR Legrest Length
To adjust & or remove the footrest length on a
Power ALR/ELR loosen the two 4.0 mm allens to
undo the footrest extension tube. Fig 8.6
The actuator is secured to the legrest by a bolt at
each end. To better access the top bolt you will
need to remove the knee pad fi rst. A 13mm open
wrench on the nut and a 13mm socket to remove
the knee pad bracket. Fig 8.7