4.23 CONNECTOR BUBBLE DETECTOR ID 1706900001
RIELE BERLIN Photometer 5010 / V4 4.23.1 23.09.2003
Function Description
The connector bubble detector connects the bubble detector with cuvette connection pcb.
Location
Optical unit
Replacement
The Photometer 5010 has to be opened. The top case has to be separated from the bottom case.
Disconnect plug P2 of cuvette board.
Loosen milled nut and remove connector bubble detector.
Installation
For installation proceed in reverse order.
See also exploded view, chapter „Dismantling“
P2
grey
blue
violet