Chapter 2 Command System RIGOL
MSO1000Z/DS1000Z Programming Guide 2-153
:TIMebase Commands
The :TIMebase commands are used to set the horizontal parameters, such as enabling the delayed sweep
and setting the horizontal timebase mode.
Command List:
:TIMebase:DELay:ENABle
:TIMebase:DELay:OFFSet
:TIMebase:DELay:SCALe
:TIMebase[:MAIN]:OFFSet
:TIMebase[:MAIN]:SCALe
:TIMebase:MODE
:TIMebase:DELay:ENABle
Syntax
:TIMebase:DELay:ENABle <bool>
:TIMebase:DELay:ENABle?
Description
Enable or disable the delayed sweep, or query the status of the delayed sweep.
Parameter
Name
Type
Range
Default
<bool>
Bool
{{1|ON}|{0|OFF}}
0|OFF
Explanation
Delayed sweep can be used to enlarge a length of waveform horizontally to view
waveform details.
Return Format
The query returns 1 or 0.
Example
:TIMebase:DELay:ENABle ON /*Enable the delayed sweep*/
:TIMebase:DELay:ENABle? /*The query returns 1*/

Table of Contents

Other manuals for Rigol DS1074Z Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1074Z Plus and is the answer not in the manual?

Rigol DS1074Z Plus Specifications

General IconGeneral
TypeTabletop
Product type-
CertificationUL 61010-1:2004; CAN/CSA-C22.2 NO. 61010-1-2004; EN 61010-1:2001; IEC 61010-1:2001
Bandwidth70 MHz
Sampling rate- MS/s
Channels quantity4 channels
Minimum time base setting (per division)5 ns
Power consumption (max)50 W
Display diagonal7 \
Non-operating altitude3000 - 15000 m
Storage temperature (T-T)-40 - 70 °C
Operating temperature (T-T)0 - 50 °C
Weight and Dimensions IconWeight and Dimensions
Depth122.4 mm
Width313.1 mm
Height160.8 mm

Related product manuals