RIGOL Chapter 2 Command System
2-8 MSO1000Z/DS1000Z Programming Guide
:CALibrate Commands
Command List:
:CALibrate:QUIT
:CALibrate:STARt
:CALibrate:QUIT
Syntax
:CALibrate:QUIT
Description
Exit the self-calibration at any time.
Related
Command
:CALibrate:STARt
:CALibrate:STARt
Syntax
:CALibrate:STARt
Description
The oscilloscope starts to execute self-calibration.
Explanation
The self-calibration operation can make the oscilloscope quickly reach its optimum
working state to obtain the most accurate measurement values.
During the self-calibration, all the channels of the oscilloscope must be disconnected
from the inputs.
The functions of most of the keys are disabled during the self-calibration. You can
send the :CALibrate:QUIT command to quit the self-calibration.

Table of Contents

Other manuals for Rigol DS1074Z Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1074Z Plus and is the answer not in the manual?

Rigol DS1074Z Plus Specifications

General IconGeneral
TypeTabletop
Product type-
CertificationUL 61010-1:2004; CAN/CSA-C22.2 NO. 61010-1-2004; EN 61010-1:2001; IEC 61010-1:2001
Bandwidth70 MHz
Sampling rate- MS/s
Channels quantity4 channels
Minimum time base setting (per division)5 ns
Power consumption (max)50 W
Display diagonal7 \
Non-operating altitude3000 - 15000 m
Storage temperature (T-T)-40 - 70 °C
Operating temperature (T-T)0 - 50 °C
Weight and Dimensions IconWeight and Dimensions
Depth122.4 mm
Width313.1 mm
Height160.8 mm

Related product manuals