Chapter 2 Command System RIGOL
MSO1000Z/DS1000Z Programming Guide 2-3
:SINGle
Syntax
:SINGle
Description
Set the oscilloscope to the single trigger mode. This command is equivalent to any of the
following two operations: pressing the SINGLE key at the front panel and sending
the :TRIGger:SWEep SINGle command.
Explanation
In the single trigger mode, the oscilloscope triggers once when the trigger conditions
are met and then stops.
When the waveform record function is enabled or during the playback of the
recorded waveform, this command is invalid.
Related
Commands
:TFORce
:RUN
:STOP
:TFORce
Syntax
:TFORce
Description
Generate a trigger signal forcefully. This command is only applicable to the normal and
single trigger modes (see the :TRIGger:SWEep command) and is equivalent to pressing
the FORCE key in the trigger control area at the front panel.

Table of Contents

Other manuals for Rigol DS1074Z Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1074Z Plus and is the answer not in the manual?

Rigol DS1074Z Plus Specifications

General IconGeneral
TypeTabletop
Product type-
CertificationUL 61010-1:2004; CAN/CSA-C22.2 NO. 61010-1-2004; EN 61010-1:2001; IEC 61010-1:2001
Bandwidth70 MHz
Sampling rate- MS/s
Channels quantity4 channels
Minimum time base setting (per division)5 ns
Power consumption (max)50 W
Display diagonal7 \
Non-operating altitude3000 - 15000 m
Storage temperature (T-T)-40 - 70 °C
Operating temperature (T-T)0 - 50 °C
Weight and Dimensions IconWeight and Dimensions
Depth122.4 mm
Width313.1 mm
Height160.8 mm

Related product manuals