Chapter 2 Command System  RIGOL 
MSO1000Z/DS1000Z Programming Guide  2-157 
:TIMebase[:MAIN]:SCALe 
:TIMebase[:MAIN]:SCALe <scale> 
:TIMebase[:MAIN]:SCALe? 
Set or query the main timebase scale. The default unit is s/div.   
YT mode: 5ns/div to 50s/div in 1-2-5 step 
Roll mode: 200ms/div to 50s/div in 1-2-5 step 
When the horizontal timebase mode is YT and the horizontal timebase is 200ms/div 
or larger (namely the "Slow Sweep" mode), this command is invalid when the 
oscilloscope is in the transition to the "Stop" state.   
The query returns the main timebase scale in scientific notation.   
:TIMebase:MAIN:SCALe 0.0002 /*Set the main timebase scale to 200μs/div*/ 
:TIMebase:MAIN:SCALe?                /*The query returns 2.0000000e-04*/ 
 
 
:TIMebase:MODE 
:TIMebase:MODE <mode> 
:TIMebase:MODE? 
Set or query the mode of the horizontal timebase.   
  MAIN: YT mode 
  XY: XY mode 
  ROLL: Roll mode 
The query returns MAIN, XY, or ROLL.   
:TIMebase:MODE XY /*Set the horizontal timebase mode to XY*/ 
:TIMebase:MODE? /*The query returns XY*/ 
:TIMebase:DELay:OFFSet 
:TIMebase:DELay:SCALe 
:TIMebase[:MAIN]:OFFSet 
:TIMebase[:MAIN]:SCALe