EasyManuals Logo

Rigol DS1074Z Plus Programming Guide

Rigol DS1074Z Plus
260 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #47 background imageLoading...
Page #47 background image
Chapter 2 Command System RIGOL
MSO1000Z/DS1000Z Programming Guide 2-31
:CURSor:AUTO:BY?
Syntax
:CURSor:AUTO:BY?
Description
Query the vertical position of cursor B in auto cursor measurement.
Explanation
The horizontal and vertical positions of the cursor are defined by the pixel
coordinate of the screen. The pixel coordinate of the screen ranges from (0,0) to
(600,400). Wherein, (0,0) is located at the left top corner of the screen and
(600,400) is located at the right bottom corner of the screen. The horizontal pixel
range is from 0 to 600 and the vertical pixel range is from 0 to 400.
In auto cursor measurement, cursor B of Y type is not required for some of the
measurement items. At this point, the query returns 4294967295.
Return
Format
The query returns an integer between 5 and 394.
:CURSor:AUTO:AXValue?
Syntax
:CURSor:AUTO:AXValue?
Description
Query the X value of cursor A in auto cursor measurement. The unit depends on the
horizontal unit currently selected.
Explanation
In auto cursor measurement, cursor A of X type is not required for some of the
measurement items. At this point, the query returns 9.9E37.
Return
Format
The query returns the X value of cursor A in scientific notation.
Example
:CURSor:AUTO:AXValue? /*The query returns -4.000000e-06*/
:CURSor:AUTO:AYValue?
Syntax
:CURSor:AUTO:AYValue?
Description
Query the Y value of cursor A in auto cursor measurement. The unit depends on the
vertical unit currently selected.
Explanation
In auto cursor measurement, cursor A of Y type is not required for some of the
measurement items. At this point, the query returns 9.9E37.
Return
Format
The query returns the Y value of cursor A in scientific notation.
Example
:CURSor:AUTO:AYValue? /*The query returns -4.000000e-01*/
:CURSor:AUTO:BXValue?
Syntax
:CURSor:AUTO:BXValue?
Description
Query the X value of cursor B in auto cursor measurement. The unit depends on the
horizontal unit currently selected.
Explanation
In auto cursor measurement, cursor B of X type is not required for some of the
measurement items. At this point, the query returns 9.9E37.
Return
Format
The query returns the X value of cursor B in scientific notation.
Example
:CURSor:AUTO:BXValue? /*The query returns -4.000000e-06*/

Table of Contents

Other manuals for Rigol DS1074Z Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1074Z Plus and is the answer not in the manual?

Rigol DS1074Z Plus Specifications

General IconGeneral
TypeTabletop
Product type-
CertificationUL 61010-1:2004; CAN/CSA-C22.2 NO. 61010-1-2004; EN 61010-1:2001; IEC 61010-1:2001
Bandwidth70 MHz
Sampling rate- MS/s
Channels quantity4 channels
Minimum time base setting (per division)5 ns
Power consumption (max)50 W
Display diagonal7 \
Non-operating altitude3000 - 15000 m
Storage temperature (T-T)-40 - 70 °C
Operating temperature (T-T)0 - 50 °C
Weight and Dimensions IconWeight and Dimensions
Depth122.4 mm
Width313.1 mm
Height160.8 mm

Related product manuals