Chapter 5 Operating
2104.6536.72 - 5.76 - E-1
Main functions
Correction > On/Off: Switches the entire frequency response correction on or off.
Correction Point 1 (2) > On/Off: Switches frequency response correction on branch 1
(or 2) on or off.
Position Point 1 (2): Entry of the threshold for the point of frequency response correc-
tion 1 (or 2). Frequency response correction is effective only if the instantaneous drive
level exceeds this threshold; setting range: 0 ... 100%
Functions, area 1
Amplitude Slope Point 1 (2) > On/Off: Switches frequency response correction on
branch 1 (or 2) for the amplitude characteristic on or off.
Amplitude Slope Amplitude 1 (2): Sets amplitude frequency response correction for
the amplitude precorrections on branch 1 (or 2); setting range: -50 ... +50%. A positive
number increases the frequency response.
Amplitude Slope Phase 1 (2): Sets phase frequency response correction for the am-
plitude precorrections on branch 1 (or 2); setting range: -50 ... +50%. A positive number
increases the frequency response.
Functions, area 2
Phase Slope Point 1 (2) > On/Off: Switches frequency response correction on branch
1 (or 2) for the phase characteristic on or off.
Phase Slope Amplitude 1 (2): Sets amplitude frequency response correction for the
phase precorrection on branch 1 (or 2); setting range: -45 ... +45%. A positive number
increases the frequency response.
Phase Slope Phase 1 (2): Sets phase frequency response correction for the phase pre-
correction on branch 1 (or 2); setting range: -45 ... +45%. A positive number increases
the frequency response.