Chapter 8 Operating Instructions
CLV 42x bar code scanner
8-6 © SICK AG · Division Auto Ident · Germany · All rights reserved 8 009 981/O078/16-08-2004
Troubleshooting
8.5.2 Malfunctions in reading mode: reading pulse errors
Prerequisite The "Device Ready" LED lights up. If not, see Table 8-3, Page 8-5.
Malfunction Possible cause Remedy
1. Pulse mode:
sensor input
(external sensor).
The CLV cannot be pulsed:
the "Laser On" LED does not light up
the scan line does not appear
Incorrect reading pulse source
parameterized on the CLV
Polarity of switching input inverted
Sensor not connected or connected to
incorrect switching input
If sensor supplied by CLV: sensor not
connected to ground potential
Photoelectric switch not aligned with the
reflector (photoelectric switch/inductive
sensor not dampened)
In the CLV Setup program: Select the
DEVICE CONFIGURATION tab and click the
READING TRIGGER PARAMETERS button.
READING TRIGGER SOURCE group: sensor
input selected?
Active high: reading pulse starts when
power applied
Connect sensor to "Sensor 1" switching
input.
Check sensor wiring (Fig. 5-4, Page 5-8).
Measure output signal on sensor.
Insert jumper between pin 5 and pin 15
Align sensor. Check functioning.
2. Pulse mode:
Reflector polling
The CLV cannot be pulsed:
the "Laser On" LED flickers but does
not light up constantly
the scan line is very faint
Incorrect reading distance (polling
reflector)
Incorrect reflector type used
Incorrect pulsing sequence (reflector is
not blocked to start the reading interval
on the CLV)
Check: reading distance of reflector in
specified reading range?
Check reflector type, Table 4-2,
Page 4-11.
Check sequence:
Read: 1. Block reflector, 2. Read bar
code, 3. Unblock reflector.
3. Pulse mode:
Serial Interface
The CLV cannot be pulsed:
the "Laser On" LED does not light up
the scan line does not appear
Incorrect reading pulse source
parameterized on the CLV
The CLV is not receiving any command
string to start the reading interval on the
data interface
In the CLV Setup program: Select the
DEVICE CONFIGURATION tab and click the
READING TRIGGER PARAMETERS button.
In the READING TRIGGER SOURCE group: is
serial interface selected?
Group SERIAL INTERFACE: is trigger type
correct?
Check data connection to host.
In the CLV Setup program: Select the
HOST INTERFACE tab.
In DATA FORMAT group: are the correct
interface type (hardware) and data
format selected?
Group: INTERFACE PROTOCOL: are the
correct start and stop characters
selected?
Use the CLV Setup program to check the
host command strings: Choose
AUXILIARY
I
NTERFACE tab. Choose MONITOR HOST
I
NTERFACE function. See Chapter 6.5.7
Monitor host interface, Page 6-21.
Download temporarily to the CLV.
4. CLV does not respond to an external
signal that ends the reading pulse
(sensor input, reflector polling, serial
interface)
Incorrect end of reading pulse
parameterized on the CLV
No sensor signal
In the CLV Setup program: Check by
choosing the DEVICE CONFIGURATION tab,
click the
READING TRIGGER PARAMETERS
button.
END OF READING INTERVAL group: is
"Generated by Trigger Source"
selected?
See Malfunction 1, Remedy
Table 8-4: Troubleshooting: reading pulse errors in reading mode