E
Eric LoveAug 2, 2025
What to do if SICK CLV 480 Barcode Reader cannot be clocked via serial interface?
- RRebecca GomezAug 2, 2025
If the SICK Barcode Reader cannot be clocked via the serial interface, and the "Sensor" LED and red scan line do not appear, verify that the reading trigger source is set to "serial interface" in the "CLV-Setup" program under the DEVICE CONFIGURATION tab, READING TRIGGER SOURCE section. Ensure you are using the correct commands for triggering, such as 21 for start and 22 for stop, or any defined single characters. Also, check the data connection to the host to confirm the CLV is receiving command strings to initiate the reading interval.