EasyManua.ls Logo

SICK DUSTHUNTER T

SICK DUSTHUNTER T
552 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GM35 Operating Instructions
Probe Model
Commissioning
56 © SICK MAIHAK GmbH • Germany · All rights reserved 8009389/07-2006
Connect the short hose piece on the differential pressure monitor securely to the purge-
air gland on the GMP measuring probe using the hose clamp.
Connect the two 6.3 mm tab receptacles of the installed signal cable to terminals 2 and
3 (NO and COM, NC contact) on the differential pressure monitor.
Fig. 30 Connecting the purge-air unit with the optional differential pressure monitor
Connecting the purge-air supply to the SR unit
Position the SR unit with measuring probe in the immediate vicinity of the flange with pipe
mounted on the measurement gas duct to connect the purge-air hose and insert the
measuring system in the duct while the purge-air supply is running.
If the angle flange, art. no. 2 017 833, was used to adjust the system, the measuring sys-
tem is usually already located at the duct flange.
Switch on the power supply to the purge-air unit briefly to test whether it is functioning
correctly, and remove any dust that may have entered the purge-air hose.
Connect the purge-air hose from the purge-air unit to the purge-air gland on the measur-
ing probe using a hose clamp. If you are using a differential pressure monitor, connect
the hose to it (see Fig. 30).
Switch on the purge-air supply.
The purge-air supply is now activated and protects the measuring system from dirt and over-
heating. For this reason, the purge-air supply must never be switched off while the measur-
ing probe is inside the measurement gas duct.
If you have not already done so, attach visible warnings to all of the switch devices that
can switch off the purge-air supply to prevent it from being deactivated inadvertently.
6.5.2 GPP measuring probes: power supply
If you are using a GPP measuring probe, the optical boundary surfaces must be heated to
the necessary operating temperature before the measuring probe is inserted in the duct:
Connect the GPP measuring probe to the power supply.
If the power supply cable of the measuring probe that is to be used for continuous mea-
suring operation cannot be connected while the measuring probe is not yet inserted in
the duct, or if it cannot be switched on, use a temporary power supply cable similar to the
one you may have used for the zero-point adjustment (Page page 54).
Purge-air hose
GMP measuring probe
Purge-air unit

Table of Contents

Related product manuals