8009441/YN36/V3-0/2015-08| SICK OPERATING INSTRUCTIONS|GM35
Subject to change without notice
START-UP
Fig. 56: Optical alignment – shown on telescope
4. Checking optical alignment
Swap the adjustment device components on the purge air fixtures around, that is, attach
the light source to the reflector flange and the telescope to the device flange.
Note the optical alignment on the telescope.
If the light point is not centered, adjust the screws of the L alignment accordingly until it
is centered.
5. Installing the SR-unit
Attach the SR-unit to the device flange:
Assemble the hinge and insert the bolt
Swivel the SR-unit closed and secure the enclosure using the quick-release fasteners.
Not aligned correctly
Aligned correctly