NOTES: 
1.  Pressing  the  <FUNCTION>  key  allows  you  to  view  status  information  on  the  sensor  that  was  verified 
in  Step  8.  Pressing  the  <FUNCTION>  key  or  the  <NEXT>  key  on  the  DISPLAY/ACTION  keypad 
repeatedly  allows  you  to  view  13  more  status  displays  on  this  particular  sensor;  using  the 
<PREVIOUS>  key  allows  you  to  go  back  and  look  at  a previously  viewed  display. 
2.  Status  displays  9  through  12  are  explained  below.  The  displays  reveal  the  following  information: 
sensor’s  sensitivity  as  programmed  at  its  node;  average  sensor  value  and  sensor  value  needed  to 
cause  an  alarm;  current  sensor  value;  and  highest  (peak)  sensor  value  to  date. 
9.  To  display  the  sensitivity  programmed  for  the  sensor,  press  the  <FUNCTION>  key  six  more  times. 
NOTE: 
If you  accidentally  press  the  <FUNCTION>  key  more  than  six  times,  you  can  go  back  one  display 
at  a time  using  the  <PREVIOUS>  key  on  the  DISPLAY/ACTION  keypad. 
l 
The  display  shows  part  of  the  custom  label  (location)  along  with  the  sensor’s  present  sensitivity. 
<Fl>  =  Change 
<F2>  =  Default 
<Enter>  = 
Save 
PRESENT  SENSITIVITY  SELECTED  =  2.5%SMOKE 
10.  Press  the  <FUNCTION>  key  again  to  display  the  sensor’s  average  and  alarm  value. 
l 
The  display  shows  sensor  location,  average  sensor  value,  and  sensor  value  needed  to  cause  an  alarm 
condition. 
I 
BLDG.  21-1  MECHANICAL  ROOM 
AVERAGE  VALUE  =  90 
/ 
ALARM  LEVEL  =  158 
I 
l 
This  display  shows  an  average  value  of  90  for  the  last  2048  values  reported  by the  sensor.  It also  shows 
that  a sensor  value  of  158  is needed  to  cause  an  alarm  condition  for  the  sensor. 
11.  Press  the  <FUNCTION>  key  again  to  display  the  sensor’s  current  value. 
l 
The  display  shows  sensor  location  and  present  sensor  value. 
21-l  MECHANICAL  ROOM 
VALUE=90  /  O%OFALARM  / 
0.0%  SMOKE 
l 
This  display  shows  the  sensor  is  currently  reporting  a  value  of  90.  It  also  shows  that  the  value  of  90 
represents  the  following:  0%  of  the  required  increase  needed  to  cause  an  alarm  and  0.0%  content  of 
smoke  in the  air  as  seen  by  the  sensor. 
12.  Press  the  <FUNCTION>  key  again  to  display  the  sensor’s  highest  (peak)  value  to  date. 
l 
The  display  shows  that  pressing  Fl  clears  a peak  value  (given  a proper  operator  access  level). 
I 
Fl  =  Clear  PEAK  value 
PEAK  =  92  /  2%  OF  ALARM  / 
0.0%  SMOKE 
66