Parameter
Command
Argument
Setting
SELF TEST
QUERIES:
Test results
Last error
TS?
ER?
FIRMWARE
QUERY
VN?
version number
CALIBRATION
QUERIES
Week of
cal.
Place ofcal.
Year
of cal.
WK?
PC?
YR?
STATISTICS
QUERIES
Standard
dev.
Maximum
Minimum
Mean
Variance
DVI
MA?
MI?
MU?
VS?
CALIBRATION
COMMANDS*
Ideal
cal.
value
Calibration
mode
CAF
CM/
0 to
5
0
1
2
3
4
5
Normal first
pass.
No magnitude first
pass.
No phase first
pass.
Second pass: phase
range-range
adjustment.
Generator calibration.
Clear calibration
data.
Forcing
w/ffreq.
DF F
Hz
Waveform
WF
/
0 sine
1
square
Year
of calibration
YR/
Week
of calibration WK/
1 to 53
Calibration
place
PC text
*These
commands
are obeyed
only when the instrument is
operating in the calibration
mode.
The
use of the calibration
commands is described in the 1255/1260
Maintenance
Manual.
JWS/1
260/2
8.13