6.2 LISTING OF EXAMPLE
PROGRAM
Command
Code Comments
roo RS 01
Recall
set-up number 1
.
00
FC
File clear, i.e. clear the
history
file.
00
MC 01
Set history file clear mode to manual. This
inhibits the
automatic file clear function and allows
the history file
to store the data from several sweeps.
00 VA +
1.0000E-01
Set the
drive signal
amplitude to 0.1V.
00 RE
Recycled
measurements This command
takes the
instrument
through the first sweep.
00 VA
+
2.0000E-01
Set
the drive signal
amplitude to 0.2 V.
00 RE
Recycled measurements.
This command takes the
instrument through the
second sweep.
00 VA +
3.0000E-01
Set
the drive signal amplitude
to 0.3V.
00
RE
Recycled measurements.
This command takes the
instrument through the third sweep.
00
VA
4-
4.Q000E-01 Set the drive signal amplitude
to 0.4V.
00 RE
Recycled measurements.
This command
takes the
instrument
through the fourth sweep.
00 VA +
5.0000E-01
Set
the
drive
signal
amplitude to 0.5V.
00 RE
Recycled
measurements. This command
takes the
instrument
through the fifth sweep.
99
#Q
Quit program.
This program
may
be
modified
as shown in Section
6.3.
JWS/1260/T
11.13