www.spacelabshealthcare.com
5-15
91390 QUBE
TROUBLESHOOTING
01030600 Power Subsystem Test failed Replace CPU PCBA.
01030601 A2D Converter failed to convert Replace CPU PCBA.
01030602 Invalid power request Replace CPU PCBA.
01030603 Temperature out of spec Replace CPU PCBA.
01030700 PCI Bridge Test failed Replace CPU PCBA.
01030701 PCI Bridge configuration not recognized Replace CPU PCBA.
01030702 PCI Bridge registers are not writable Replace CPU PCBA.
01030703 PCI Bridge revision not valid for clinical use Replace CPU PCBA.
01030800 Ethernet Test failed Replace CPU PCBA.
01030801 Ethernet Setup failed Replace CPU PCBA.
01030802 Ethernet Transmit failed Replace CPU PCBA.
01030A00 Video Test failed Replace CPU PCBA.
01030A01 Invalid display type Replace CPU PCBA.
01030A02 Invalid display size Replace CPU PCBA.
01030A03 Video configuration not recognized Replace CPU PCBA.
01030A04 Video DRAM failure Replace CPU PCBA.
01030A05 Video could not detect H-sync signal Replace CPU PCBA.
01030A06 Video could not detect blue video signal Replace CPU PCBA.
01030C00 OS error while diagnosing KBD Reboot and retest. If problem
persists, replace CPU PCBA.
01030C01 Cannot access KBD device registers Replace CPU PCBA.
01030C02 KBD did not respond to command Replace keyboard. Replace CPU
PCBA.
01030C03 KBD failed self test Replace keyboard. Replace CPU
PCBA.
01030C04 KBD failed interface test Replace keyboard. Replace CPU
PCBA.
01030D00 Cannot access AUDIO device registers Replace CPU PCBA.
01030D01 Could not open audio device Reboot and retest. If problem
persists, replace CPU PCBA.
01030D02 Audio device did not respond correctly Replace CPU PCBA.
01030D03 Audio Loopback test failed Replace CPU PCBA.
Table 5-1 Diagnostic Failure Messages (Continued)
Error Code Diagnostics Failure Message Suggested Action