TROUBLESHOOTING
System Troubleshooting
This section describes troubleshooting procedures. You can use
these procedures with or separately from the diagnostics to isolate a
failure.
Suggested Tools/Test Equipment
• Anti-static mat with wrist strap
• #1 and #2 Phillips-head screwdriver
• 3/16-inch nutdriver standard flat screwdriver
•Multimeter
01030E00 NVRAM Test failed Replace CPU PCBA.
01030E01 NVRAM Clock not running Replace CPU PCBA.
01030E02 NVRAM checksum error Zero the NVRAM and reboot.
Re-program all items in the
Biomed and Clinical menus
(Spacelabs Healthcare Field
Service Engineers only). If
problem persists, replace CPU
PCBA.
01030E03 NVRAM read/write memory test failed Replace CPU PCBA.