A PPENDIX A — ELECTROMAGNETIC
C OMPATIBILITY
Mitigation
• Move AC line connection to another AC outlet/power phase.
• Position power cable and power supply away from other devices
and their power and data cables.
• Fold excess cable length in a noninductive bundle. Do not coil or
roll long cables.
• Change power supplies.
• Power the monitor/module huousing through an isolation
transformer.
Rated maximum
output power of
transmitter (watts)
Separation distance according to frequency of transmitter
(meters)
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
0.01 0.02 0.02 0.04
0.1 0.06 0.06 0.1
1 0.2 0.2 0.4
10 0.6 0.6 1.1
100 1.8 1.8 3.5
Notes:
• At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
• These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.