www.spacelabshealthcare.com
4-7
91390 QUBE
M AINTENANCE
An alternate method is to connect the cable and turn the monitor
ON and then OFF. Each color of the embedded alarm light
momentarily illuminates. An external alarm light also briefly
illuminates.
Ethernet (Wired)
1 Attach a 10/100BaseT cable from an active network onto the rear
panel connector.
2 Verify that the green link light on the connector is ON.
Assembly/Disassembly Procedures
Note:
The external power supply is not designed for disassembly.
Required Tools and Parts
Anti-static mat with wrist strap
#1 and #2 Phillips-head screwdriver
3/16-inch nut driver
Tool, antenna extraction, wireless card, P/N 003-0286-00
XPREZZON™ and qube™ System Administration Manual, P/N
070-2380-xx
Setup for Disassembly
The battery switch has symbols to show when to turn the switch on
and off. Turn the switch off for shipping and service (refer to
Figure 4-1). Turn the switch to on for regular operation (refer to
Figure 4-2).
Figure 4-1 Battery Switch Off Symbol
Caution:
Before you begin any disassembly procedures, power OFF the monitor,
disconnect the AC cord from the AC power receptacle, and remove the
batteries.