www.spacelabshealthcare.com
4-21
91390 QUBE
M AINTENANCE
Mon-Dock PCB
1 Perform Step 1 on page 4-18 through Step 9 on page 4-19 to
access the Main Chassis assembly.
2 With the Chassis assembly face down, remove the seven screws
shown in Figure 4-15 on page 4-21 for option A. Remove the five
screws for option B or C.
Figure 4-15 Docking PCB
3 Replace the old PCB. Attach it again to the bottom of the chassis
assembly with the original screws.
4 Perform the retest procedure. Refer to Functional Tests on
page 4-5.