EasyManua.ls Logo

Spacelabs qube 91390 - To Change Batteries Without Interruptions While Operating on Battery Power; Battery Power Level

Spacelabs qube 91390
160 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
www.spacelabshealthcare.com
1-15
91390 QUBE
I NTRODUCTION
One battery charges in approximately two hours, and two
batteries charge in approximately four hours.
The option B monitor operates for up to four hours with two fully
charged batteries and standard configuration. With a single fully
charged battery and standard configuration, the monitor operates
for up to two hours. The battery is not removable from the monitor.
The option C monitor operates for up to two hours with a single
fully-charged batteries and standard configuration. The battery is
not removable from the monitor.
Note:
For options B and C monitors, only one battery icon shows on the
main screen.
To change batteries without interruptions while
operating on battery power
1 If operating on top battery (battery A), remove lower battery
(battery B).
2 Install new battery (battery C) into the lower battery slot.
3 To make sure that the operation of the new battery is correct,
look at the battery indicator lights (Refer to Table 1-5 on page 1-
15).
4 Remove battery A. Replace it with another new battery.
Battery Power Level
Extended use of the batteries depletes the battery power. To
prevent patient data loss, the qube monitor watches the battery
power level that remains. As power gets low, the monitor warns the
user about low battery power and inhibits the use of the internal
recorder to extend run time.
Caution:
The qube monitor and smart rechargeable Li-Ion batteries and cells do
not overheat, they have internal protection devices. Overexposure to heat
will activate one of these devices and may permanently disable the
battery. If this happens, contact Spacelabs Healthcare to purchase a
replacement battery.
Table 1-5 Battery power level messages and tones
A green color on the battery icon shows the current charge level of the battery.
A yellow or red color shows that the battery power is at a critically low level. If the battery power is critically
low, recharge the battery.

Table of Contents