www.spacelabshealthcare.com
4-35
91390 QUBE
M AINTENANCE
Cleaning
To clean the case, wash it with mild soap and water or use Plast-N-
Glas cleaner. To clean the electronic connectors and contacts, use
TF solvent as necessary.
Refer to the XPREZZON and qube Bedside Monitoring Operations
Manual (P/N 070-2112-xx) for more information.
Caution:
• Do not autoclave.
• Never use solvents, acetone, abrasive cleaning agents, or abrasive
cleaning pads.
• Only use approved cleaning agents, including 70% alcohol, soap and
water, green soap, or 10% bleach solution.
• Do not directly spray liquids into the recorder, module, or battery
compartments.