M AINTENANCE
Replace and Align the Module Door
1 Remove the chassis. Refer to Complete Main Chassis Removal on
page 4-17.
2 To remove the door, remove the two screws fastening the hinge
to the wall of the module compartment assembly. This frees the
door assembly. When you install a new door, do not tighten the
two screws until you completely secure the chassis assembly to
the rear housing.
3 After the chassis and rear housing are reassembled, position the
door so that it freely opens and closes, and then tighten the two
screws.
Figure 4-20 Module Door