www.spacelabshealthcare.com
4-31
91390 QUBE
M AINTENANCE
Docking Station PCBA Replacement Procedures for
Option A
Required Tools and Parts
• Antistatic mat with wrist strap
• #1 and #2 Phillips-head screwdriver
• 3/16-inch nut driver
• Docking Station PCBA (P/N 670-1558-xx)
Docking Station
Figure 4-27 Qube Docking Station
1 Disconnect all cables from the docking station.
2 Turn the docking station over. Remove the six screws from the
docking station cover plate. Refer to Figure 4-28 on page 4-32.
Caution:
Before you begin any disassembly procedures, disconnect the AC cord
from the AC power receptacle. Make sure to use the proper procedures to
handle electrostatic-sensitive devices.