EasyManua.ls Logo

Spacelabs qube 91390 - Battery Switch PCBA for Options B and C

Spacelabs qube 91390
160 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
www.spacelabshealthcare.com
3-7
91390 QUBE
T HEORY
Battery Switch PCBA for Options B and C
Figure 3-5 Battery Switch PCBA for Options B and C
The batteries ship pre-installed for qube options B and C and require
unit disassembly for replacement. The batteries used on these
options are the same type of Li-Ion smart battery used on option A,
but have a lower capacity and are physically shorter. The 670-1658-
xx battery board used on options B and C includes a small micro
switch mounted on the bottom left of the monitor (refer to
Figure 3-
6).
BACKLIGHT
J40
J9
INPUT
SELECT
DC_IN
V BATT 1
V BATT 2
3.3VAO
“ALWAYS
ON”
BATTERY
CHARGER
CPU
“ALWAYS ON”
MCU
CONTROL
BAT1_CH_LED
MAINS_LED
PWR_ON_LED
SW_PWR
RKEY
FP LEDs & IR
J50
BAT2_CH_LED
+3.3VAO
FET
SWITCHES
CH_LED2
CH_LED1
+3.3V
DC_IN
+5V / +3.3V
BUCK
-12V INVERT
+12V/+12V_REC
BUCK-BOOST
+5V
EN_3V
EN_5V
EN_18V
EN_12V
VCOM
EN_INV
+3.3V
-12V
+12VREC
+12V
EN_5V
EN_3V
EN_INV
EN_12V
VCOM
FAN
FAN_CTL
VCOM
VCOM
VCOM
+12V
+3.3VAO
J8
FAN_CTL
SMART
BATTERY 2
J9
P401 P402
BL_PWM
+12V
BL_PWM
+5V
TO LCD/ B EZZEL ASSY
TO FPGA PLD
TO LCD/ B EZZEL ASSY
+3.3VAO
+3.3VAO
BA TTERY CONNECTOR PCB A
670-1658-xx
DC_IN
PWR_RX_TX
TO
MAIN
CPU
ANY_PWR_IN
TO SYSTEM
FAILURE ALARM
PFAIL
WD_PWR
TO SYSTEM
FAILURE ALARM
WD_PWR
WD_MAIN
SYSTEM FAILURE ALARM
ANY_PWR_IN
FROM MAIN CPU
FROM P.S. “ALWAYS ON” CPU
PIEZOELECTRIC
BUZZER
FET
+5v
REG
ONE-
SHOT
ONE-
SHOT
SUPERCAP
V_ALARM
DUAL-
FREQUENCY
OSCILLATOR
670-1573-02 and later
CPU PCBA only
BATTERY
DISCONNECT
SMART
BATTERY 1
BATTERY
1 SENSE
BATTERY
2 SENSE
SW1
TO MUX
TO MUX
ADC
+3.3V
+5V
+12V
DC_IN
VCOM
-12V
BATT 1 SENSE
BATT 2 SENSE
MUX
ADC
SELECT
VCOM
DC_IN
P/O CPU
PCBA
670-1573-xx
DC_IN
TO/FROM
MON-DOCK PCBA

Table of Contents

Related product manuals