5-1
91390 QUBE
Troubleshooting
Overview
The first several sections describe the available diagnostics features
and how to use them. Next is a section which lists diagnostics failure
messages and the suggested corrective actions. The last section
gives specific troubleshooting steps which can be used to isolate
failures.
The troubleshooting procedures in this chapter isolate equipment
problems to a Field Replaceable Unit (FRU).
The following items are field replaceable:
PCBA, CPU, P/N 670-1573-xx (all options)
PCBA, Interconnect, P/N 670-1481-xx (options A and B)
PCBA, Interconnect, P/N 670-1644-xx (option C)
PCBA, Monitor Dock, P/N 670-1521-xx (option A)
PCBA, Monitor Dock, P/N 670-1647-xx (options B and C)
PCBA, Connector, P/N 670-1522-xx (option A)
PCBA, Connector, P/N 670-1645-xx (options B and C)
PCBA, Battery Connector, P/N 670-1555-xx (option A)
PCBA, Battery Connector, P/N 670-1658-xx (options B and C)
External DC Power Supply, P/N 119-0552-xx (all options)
Pod interface, P/N 670-1550-xx (options A and B)
Pod interface, P/N 670-1559-xx (option C)
Caution:
Observe precautions for handling electrostatic-sensitive devices!