www.spacelabshealthcare.com
4-33
91390 QUBE
M AINTENANCE
Figure 4-30 Latch Docking Station
7 Remove the six (6) connector standoff screws from the back of
the docking station with a nut driver. Refer to Figure 4-31.
Figure 4-31 Docking Station Connectors
8 Turn the docking station back over. Carefully slide out the
docking station PCBA. Do not damage the connectors.
9 Install a new docking station PCBA in the reverse order.
10 Perform functional tests in accordance with the monitor
functional tests. Refer to Figure 4-30.
Replace the Docking Station Latch
1 Disconnect all cables from the docking station.
2 Turn the docking station over. Remove the six screws from the
docking station cover plate. Refer to Figure 4-27 on page 4-31.
3 Remove the docking station cover plate. Set it aside.
4 Remove the connection J70 (LED wiring assembly) from the
docking station PCBA. Refer to Figure 4-32.