5-7
91390 QUBE
TROUBLESHOOTING
Figure 5-5 Boot Parameters Menu with Factory Default Settings
(v3.03.00 and higher)
Extended Diagnostics
Extended diagnostic tests, like the power-ON diagnostics, can be
used to troubleshoot and isolate many system failures. The types of
tests and features available in the extended diagnostics are:
Power-ON diagnostics.
Interactive tests and read/write memory tests that are not
appropriate during power-ON diagnostics.
•Touchscreen calibration and data dump utilities.
System data dump and system reset utilities.
The diagnostic menus allow most of these tests to be run
individually or all at once. If Loop mode is activated, a test can be
executed in a continuous loop. If Halt On Error mode is activated,
the looping stops when a diagnostic failure is detected. To avoid
false failures, do not use the touchscreen, mouse, or keyboard while
the diagnostic tests execute.
Any error that occurs during one of these tests is logged in the non-
volatile memory (NVRAM). Refer to Error Log on page 5-11 to
retrieve the log.
For detailed information on extended diagnostics and how to run
them, Refer to Diagnostic Menus on page 5-8.