4-8
91390
QUBE
M AINTENANCE
Figure 4-2 Battery Switch On Symbol
For Option A
1 Remove the external power supply cable and battery or batteries,
if present.
2 Remove the patient parameter module.
For Options B and C
1 Remove the external power supply cable and turn the battery
switch to off.
2 Remove the patient parameter module.
Remove the Bezel Assembly
1 Place the monitor face-down on a stable and padded surface.
2 Remove the eight finishing plugs from the rear enclosure as
shown in Figure 4-3 for option A.
3 Remove the 10 screws from the rear enclosure, as shown in
Figure 4-3.