5-28
91390
QUBE
TROUBLESHOOTING
Ethernet/Network
Check that the green LED on the network connector is lit showing
“Link”, and blinks with network activity. If the link light is lit, the
Ethernet connection between the monitor and the network switch is
functional. Network communication issues related to monitoring
functions can be due to incorrect network setup configuration in the
monitor, or network infrastructure issues.
1 Check the network setup in the Biomed menu.
2 Check that the correct Monitor ID, Monitor Name, IP address, IP
subnet, IP Gateway, Network Number, Network Size, and DSCP
values are correct.
3 Contact the responsible IT department for network infrastructure
issues.
External Alarm Relay/Nurse Alert for Option A
1 Check that the Alarm Relay Setup items under the Alarms tab in
the Biomed menu are configured correctly.
Figure 5-20 Alarm Relay Setup
2 Check that the alarm cable is securely connected to the 15-pin
connector and the remote device.
3 Check that pin 9 on the connector has +12 VDC.
4 Check that the custom-design remote alarm device is functional.
5 Check and replace in order:
1 CPU PCBA
2 Connector PCBA

Table of Contents