www.spacelabshealthcare.com
4-18
91390
QUBE
M AINTENANCE
Figure 4-12 PCB Standoffs
7 Open module bay door to access and disconnect the 175-1860-xx
Pod interface cable from the Pod Interface Board for options A or
B.
8 Remove the DSB and docking connector covers as shown in
Figure 4-26 on page 4-30 for option C.
9 Remove the printer, if present in options A or B.
10 With the module bay door open, pull the main chassis from the
rear enclosure.
Note:
The DSB connector must clear the rear enclosure. To allow the
DSB connector to clear, gently flex the rear enclosure.
Reinstall the Chassis
1 With the module bay door open, slide the chassis assembly into
the rear enclosure for options A or B.
2 Attach the 175-1860-xx Pod interface cable to the Pod Interface
Board again.
3 Install the four chassis screws again for option A. Install the three
chassis screws again for options B or C.