TROUBLESHOOTING
Pod does not sign on
If a patient module installed in the module bay of the monitor signs
on, but the Pod does not:
1 inspect the Pod Interface Cable and its connections between the
Pod Interface and Interconnect PCBAs.
2 Replace parts in this order:
1 Pod Interface Cable
2 Pod Interface PCBA
3 Interconnect PCBA
Module signs on but parameters do not function
If a module signs on when installed in the module bay of the monitor,
but parameters such as ECG do not function:
1 Check that the +12vdc and -12vdc voltages are present on the Pod
Interface PCBA. Refer to Figure 5-19.
2 If either of the voltages are missing, replace in order:
a Interconnect PCBA
b CPU PCBA
3 If the +12vdc and -12vdc voltages are present, replace the
Interconnect PCBA.
Ethernet
1 The green link LED lights up when the RJ45 cable is connected
between the monitor and the network switch. The green link LED
blinks, which indicates network activity.
2 If the green LED is not lit, inspect the cabling between the
monitor and switch. Make sure that the Ethernet switch functions
correctly. If the cabling and network infrastructure are good,
replace the CPU PCBA.
3 If the green LED is lit, the Ethernet hardware in the monitor
functions properly. If the monitor does not communicate with
other monitors on the network, the problem is either:
•in Network Setup menu
•in the Biomed menu
• a configuration issue in the network switch or infrastructure