2-2
91390
QUBE
S ETUP
Notes:
Before you touch electrostatic-sensitive electronic
components, follow proper anti-static procedures, including
the use of an ESD wrist band and mat. An electrostatic
discharge from your fingers can permanently damage
electronic components and cause latent failures.
All static-sensitive electronic components are packaged in
static-shielding bags. Keep the bag in case you must
repackage the component to store it or return it to Spacelabs
Healthcare for any reason.
For options B and C, the battery switch must be set to ON
prior to use.
Display Assembly
Display features:
LCD TFT Display, 12.1" with backlight control
5 Wire Resistive Touch screen
RED, YELLOW, and CYAN LED bars: options A and B
Remote control IR receiver: option A
Power and Battery status indicators (LEDs)
•Power switch
Warning:
The qube monitor and smart rechargeable Li-Ion batteries and cells do
not overheat, they have internal protection devices. Overexposure to
heat will activate one of these devices and may permanently disable
the battery. If this happens, contact Spacelabs Healthcare to purchase
a replacement battery.
Failure to properly configure this device prior to network connection
may alter network time or corrupt patient data.
Do not mount the monitor or docking station directly above the
patient.
Caution:
If liquid spills onto the docking station, make sure that the docking
station is checked by service staff before you use it again.
Observe precautions for handling electrostatic-sensitive devices!