www.spacelabshealthcare.com
4-19
91390 QUBE
M AINTENANCE
Note:
For option A, install the hidden screw on the lower-right side of
the chassis through printer slot and hold the screw in place.
4 Install the connector PCB Standoffs again.
5 Place the monitor on its rear. Reinstall the five screws on the
bottom of the monitor.
6 Install the wireless card again, if applicable.
7 Attach antenna cables to the wireless card again, if applicable
8 Install the bezel again. Refer to Step Step 15 on page 4-13 through
Step 22 on page 4-14.
9 Perform the retest procedure. Refer to Functional Tests on
page 4-5.
SDLC Interface PCB
1 Remove the Bezel Assembly by perform Step 1 on page 4-8
though Step 7 on page 4-10.
2 Perform Step 2 on page 4-15 through Step 7 on page 4-18 to
remove the Main Chassis assembly.
3 Remove the CPU. Refer to CPU PCB on page 4-11
4 Remove the two screws shown in Figure 4-13 on page 4-19.
Figure 4-13 SDLC Interface