EN-8
Rear Panel
The console rear panel provides ports for connecting the 1488 HD Camera to
viewing and recording equipment, such as video monitors and Stryker device
control consoles.
See the “Product Description” section for the dierent console models that
are available. The 1488 HD 3-Chip Camera Control Unit with DVI Fiber Output
(1488010001) is shown below.
1 2 53 4 6 7 8 9
10
11
1. Device Control Port Connects to a Stryker device control console
(such as the SDC3 or SIDNE®) to enable voice
operation and/or graphic tablet control
2. Remote Out 1 Connects to a video accessory remote input
3. Remote Out 2 Connects to a video accessory remote input
4. S-Video Out Analog video output
5. DVI Out 1 Digital video output
6. DVI Out 2 Digital video output
7. Light Source Port Connects to Stryker L9000 Light Source
(0220210000)
8. AC Power Inlet Connects to AC mains with separable power
cord
9. Fuse Panel Contains two 1.6A 250V (slow blow, high
breaking capacity 1500A) fuses
10. Equipotential
Ground Plug
Connects to a potential equalization conductor.
The resulting medical electrical system shall
follow all applicable IEC 60601-1 requirements.
11. Fiber Outputs
(optical)
DVI output for connection to Lucent connector
bers (optional: 1488010001)