Performing the White-Balance Test
Before each surgical procedure, perform the white-balance test to adjust the
cameras perception of white so it can display other colors correctly.
1. Ensure that a scope and light source are attached to the camera, and that
the camera, light source and monitor are powered on.
2. Point the scope tip at several stacked white gauze pads, a white
laparoscopic sponge, or any clean white surface.
3. Look at the monitor and make sure there is no visible glare o of the
white surface of the image.
4. Press and hold the camera head W button (or WB” on the touchscreen)
until the monitor displays the message WHITE BALANCE IN PROGRESS.
5. Continue pointing the scope at the white surface until the video monitor
displays the message WHITE BALANCE COMPLETE. The video picture
may change color. If you cannot achieve an acceptable white balance,
refer to the Troubleshooting section.
Advanced Features
The 1488 HD Camera has additional features that are not detailed in this
guide:
• Video image settings
• Button programming
• Touchscreen language settings
• L9000 light source “Run/Standby controls
• Other system settings
These advanced features require in-depth knowledge of the device and
should be performed only by trained personnel. For access to advanced
features, contact a Stryker representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker 1488 and is the answer not in the manual?

Stryker 1488 Specifications

General IconGeneral
Resolution1920 x 1080
Sensor TypeCMOS
White BalanceAutomatic/Manual
OutputHDMI, DVI, HD-SDI
Camera Head TechnologyThree-chip

Related product manuals